00-f5 Title and Logo 00-f6
 guide Contents
 siteinfo Site Info


Chapter 17

17-01
Introduction

17-02
Field Perturbations

Local Inhomogeneities
Susceptibility
17-03
Radiofrequency and Gradient Artifacts

Slice Profile
Multiple Spin-Echo
Line Artifacts
17-04
Motion and Flow Artifacts

Respiratory Motion
Cardiac Motion
Flow Artifacts
17-05
Signal Processing and
Signal Mapping

Chemical-Shift
Black Boundary
Truncation
Aliasing
Quadrature Artifacts
k-Space Artifacts


17-04 Motion and Flow Artifacts

17-04-01 Respiratory and Cardiac Motion

Motion artifacts are the most frequently observed artifacts in MR imaging and have severely hampered the use of MR imaging for abdominal studies. They re­qui­re that ECG triggering be used for thoracic studies (Figure 17-07).


Figure 17-07:
Ghost images resulting from respiratory and cardiac motion with both sets of ar­ti­facts being oriented along the phase en­cod­ing gradient. The respiratory motion produces a number of distinct ghost ima­ges of the chest wall, while the cardiac mo­tion results in the column of noise as­so­cia­ted with the heart.


Motion can lead to blurring and ghost artifacts. Blurring of anatomical stru­ctu­res and interfaces results through averaging of moving structures (pseudo doub­le exposure). This may obscure small lesions. Ghost images are partial co­pies of the parent image appearing at a different location; they are mostly caus­ed by pul­sa­ti­le flow.

Motion can be divided into two basic categories:

spaceholder 600 motion occurring between acquisition of different lines in the study;
spaceholder 600 motion occurring between excitation and data acquisition.


spaceholder blue The first category can be totally eliminated by physiological gating so that the data acquisition and motion are synchronous. This is widely used in cardiac studies and allows excellent, reproducible images of the heart cycle to be ob­tain­ed.

Respiratory gating has also been tried, but the much lower frequency of the respiratory cycle results in very long scan times. A number of other techniques monitor the respiratory cycle and select the phase-encoding steps in an order that minimizes the resulting artifact. The ROPE (Respiratory Ordered Phase Encoding) technique removes the periodicity of breathing upon k-space [⇒ Bai­les].

A technique developed later follows the normal sequence with a second ac­qui­si­tion which has no phase-encoding. This second echo, also called navigator echo, provides an indication of the amount of motion and can be used as the basis for a postprocessing procedure [⇒ Ehman]. All these techniques suffer from the fun­da­men­tal drawback that they assume bulk motion, i.e., everything is moving in the same direction at the same speed, which is not true in the abdomen. However, such techniques can improve image quality.

An alternative approach is to make the scan time short with respect to the re­spi­ra­to­ry cycle and hence limit the amount of motion. The ultimate example of this is echo-planar scanning where the image is formed from a single acquisition [⇒ Rzedzian].

By using fast imaging techniques with acquisition times of a few seconds or less in conjunction with breath holding excellent images with few motion ar­ti­facts can be obtained.

spaceholder blue Artifacts resulting from the second type of motion can be reduced by using motion-compensated gradients. The gradients in standard imaging sequences produce additional phase shifts for moving samples. Since the amount of motion will not be the same for each line of the scan (unless gating is used), a smearing of the signal in the phase-encoding direction will result.

By using a modified form for the read and slice gradients, one can remove this additional phase shift and hence the artifact (MAST — Motion Artifact Sup­pres­sion Technique) [⇒ Pattany]. The drawback is that the minimum echo time will be somewhat longer for such sequences.


17-04-02 Flow Artifacts

The origin of flow artifacts is very similar to that of motion artifacts, namely that the blood and CSF flow can be pulsatile. Thus, different flow velocities will be pre­sent in different lines of the scan. The read and slice gradients induce a phase shift for flowing material, resulting in a range of phase shifts being produced in the course of a study [⇒ van Dijk]. The resulting artifact can take the form of a ge­ne­ral smearing or a number of distinct artifacts in the phase-encoding direction (Figure 17-08).


Figure 17-08:
The misregistration of flowing spins leads to the depicting of blood outside the vessel lumen: flow artifact.


The solutions are the same as for motion artifacts: the use of ECG gating to en­su­re that the same flow velocity is always observed and (or) the use of com­pen­sa­ted gradients to null the phase shift from flowing material.

Flow artifacts are particularly severe in gradient-echo images since in a 2D study, the flowing spins will not have experienced the preceding pulses and will therefore be fully recovered. This leads to a very strong signal from the flowing blood, and therefore to severe artifacts in the absence of gating and motion com­pen­sa­tion (Figure 17-09).


Figure 17-09:
Gradient-echo image of a neck with the pha­se en­cod­ing gradient oriented ver­ti­cal­ly. Flow artifacts are observed as a column associated with arterial vessels. Venous flow also produces artifacts, but at a much lower level since this kind of flow is less pul­sa­ti­le.


Where flow of blood or CSF degrades the diagnostic quality of the MR image, artifacts should be eliminated. This can be achieved by presaturation.

Here, additional RF pulses are applied which saturate spins outside the ima­ged region. Thus, blood flowing into this region is saturated, which causes a re­duc­tion in blood signal intensity and in flow artifact. Usually this is performed on both sides of the imaged slice because blood can enter the slice from either di­rec­tion. Figure 17-10 is a theoretical example of parallel presaturation.


Figure 17-10:
Example of parallel presaturation. Presaturation pulses excite spins outside the slice to be imaged. Thus, flowing blood ar­ri­ves already saturated when the spins in the slice of interest are exposed to the ex­ci­ta­tion pulse. The blood signal is sup­pres­sed and does not give rise to artifacts. The arrows indicate the direction of flow.

spaceholder 600 spaceholder 600

LogoTop
LogoBottom
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4
space
00-f7
space
00-f1
space
00-f2
space
00-f3
space
00-f4